12.2 External Signal Input (External Trigger Input)
219
Chapter 12 Ext
ernal Cont
rol
12
Triggering can be controlled by applying a signal from an external trigger source (p.
127). Furthermore, an event mark can be added by an input signal (p. 142).
This allows synchronous operation of multiple Memory HiCorders by parallel trigger-
ing (p. 226).
12.2 External Signal Input (External Trigger Input)
External signal
input (EXT.TRIG)
Trigger event occurs.
External
Trigger Input Signals
Voltage range
HIGH level: 2.5 to 5.0 V, LOW level: 0 to 0.8 V
Pulse width
• Filter Off:
HIGH level: 1 ms or greater, LOW level: 2
μ
s or greater
• Filter On:
HIGH level: 2.5 ms or greater, LOW level: 2.5 ms or greater
Maximum input voltage
0 to 10 V DC
100 pF
3.3 k
Ω
5 V
5 V
GND
EXT.TRIG
HIGH
2.5 to 5.0 V
LOW
0 to 0.8 V
[
↓
]
[
↓
]
1 ms or greater (Filter Off)
2.5 ms or greater (Filter On)
2
μ
s or greater (Filter Off)
2.5 ms or greater (Filter On)
1 k
Ω
When using an external trigger to synchronize multiple instruments, make
sure you set
[Filter]
of the external trigger input to
[Off]
.
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......