![Hioki MR8880-20 Instruction Manual Download Page 100](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332100.webp)
4.6 Setting the Screen Display
90
When High-speed Function (Interval Setting) and Real-time Function
Change the time per division of the horizontal axis
1
2
Select from the setting options of
[Time/div].
The setting can also be
changed on the Waveform
screen. ([Wave+Set] Display)
Setting options (
∗
. default setting)
If the recording interval is changed after measure-
ment, the selection range for the horizontal axis
changes, but the horizontal axis settings which actual-
ly allow you to expand or compress the horizontal axis
are those of the selection range that corresponds to
the recording interval used during measurement.
[Highspeed]
10
μ
s, 20
μ
s, 50
μ
s, 100
μ
s
∗
, 200
μ
s,
500
μ
s, 10ms, 20ms, 50ms, 100ms,
200ms, 500ms, 1s, 2s, 5s, 10s, 30s,
1min, 2min, 5min, 10min [/div]
[Realtime]
10ms, 20ms, 50ms, 100ms, 200ms,
500ms, 1s
∗
, 2s, 5s, 10s, 30s, 1min,
2min, 5min, 10min, 30min, 1h, 2h, 5h,
10h, 12h, 1day [/div]
1 div
(Open/confirm the setting information.)
Selection Range of Horizontal Axis
The selection range of the horizontal axis differs depending on the function and [Enve-
lope] function.
• High-speed Function: A value equivalent to the setting value of [Interval] multiplied
by less than 10 cannot be set.
(Example: When the recording interval is 1 ms, the horizontal axis can be set from 10
ms.)
• Real-time Function when [Envelope] is [Off]:
A value equivalent to the setting value of [Interval] multiplied by less than 2 cannot be
set.
(Example: When the recording interval is 10 ms, the horizontal axis can be set from
20 ms.)
• Real-time Function when [Envelope] is [Off]:
A value equivalent to the setting value of [Interval] multiplied by less than 100 cannot
be set.
(Example: When the recording interval is 10 ms, the horizontal axis can be set from 1
s.)
Select the
[Setting]
screen.
Values of 1 s/div and above are displayed during real-time saving. If the
setting value is less than 1 s/div, it is changed to 1 s/div simultaneously
with the start of measurement.
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......