![Hioki MR8880-20 Instruction Manual Download Page 94](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332094.webp)
4.4 Using Real-time Function (Fluctuation Recording)
84
The maximum and minimum values within the
recording interval are calculated from data oversam-
pled at maximum sampling speed (1 MS/s), and
recorded for each recording interval. Therefore,
recording without missing the peak of fluctuation will
be possible even if a long recording interval is set.
(*): Sampling is the acquiring of the data to be saved
to the internal buffer memory. Oversampling is the acquiring of data at a higher
speed than when data is acquired to the internal buffer memory (sampling). Over-
sampled data is not saved.
Retaining Records of Maximum and Minimum Values (Envelope)
MAX
MIN
Sampling interval
1 M
1
2
(Open/confirm the setting information.)
Select from the setting options of
[Envelope]
.
Setting options (
∗
. default setting)
On
Records the maximum value and minimum
values within the sampling interval.
Off
∗
Records instantaneous data of each sampling
interval.
Select the
[Setting]
screen.
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......