12.2 External Signal Input (External Trigger Input)
220
Signal Input Procedure
1
Connect the cables for the corresponding
external input signals to the EXT.TRIG and
GND terminals.
2
Press the
SET
key to open the
[System]
screen.
3
Select an item from
[External Trig In]
of
external control.
4
Short-circuit the EXT.TRIG terminal and
GND, or leave the terminals open-circuited,
and input a HIGH level (2.5 to 5.0 V) or LOW
level (0 to 0.8 V) pulse wave or rectangular
wave to the EXT.TRIG terminal.
Trigger
Applies a trigger with an input signal.
Event
Adds an event mark with an input signal.
See p. 217 for the connection
method.
Select either rising or falling.