12.1 Connecting to the External Control Terminals
217
Chapter 12 Ext
ernal Cont
rol
12
The external control terminals on the Memory HiCorder support trigger signal input
and output.
External Control
Chapter 12
1
START/IN1
Input signals for controlling the operation of the
instrument.
(Start, stop, abort, print)
Input
2
STOP/IN2
3
GND
(common with instrument ground)
4
ALM/OUT
Output signals for instrument status.
(Printing, saving, instrument alarm)
Output
5
TRIG.OUT
Output signal when trigger applied.
Output
6
EXT.TRIG
(External trigger input)
Input external signals.
Input
External control
terminals
(Right side)
12.1
Connecting to the External Control
Terminals
To avoid electrical hazards and damage to the instrument, do not
apply voltage exceeding the rated maximum to the external con-
trol terminals.
I/O terminals
Maximum input voltage
Input
EXT.TRIG
0 to 10 V DC
START/IN1
STOP/IN2
Output
ALM/OUT
0 to 50 V DC
TRIG.OUT
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......