Note
One way to overcome exceeding the maximum external sample rate of the card
of choice might be to use the “External Clock Divider”. The maximum external
clock rate of the GEN DAQ mainframe is 2 MS/s. Using an “External Clock
Divider” of 4 would allows the 1 MS/s card to run an external clock rate of
2 MS/s. It does not increase the measurement resolution, but it does allow a
measurement at higher external clock rates.
H.1.4
Perception Display settings explained
The Perception displays are capable of handling external time base recordings
in different ways. The setup is controlled by using the display
’
s X
-
Annotation
Scaling settings.
These settings can be found by right clicking the mouse in the display of choice
and by selecting “Display Setup” at the bottom of the menu.
Figure H.4:
Perception
-
Display setup
A
Display setup
GEN3t
896
I3995-3.1 en HBM: public