11.3.1
Connecting the Master/Slave Synchronization connector
With the fiber optic cable, connect the Master/Slave Synchronization connector
of two mainframes.
In this setup, one mainframe must be set to Master while the other mainframe
must then be set to Slave. In this setup, automatically extended synchronization
is used.
POWER
1 MW 250Vpk
1 MW 250Vpk
1
2
3
4
5
6
7
8
1 MW 250Vpk
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
POWER
1 MW 250Vpk
1 MW 250Vpk
1
2
3
4
5
6
7
8
1 MW 250Vpk
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
B
A
C
Figure 11.5:
Connecting the Master/Slave
A
Master mainframe
B
Slave mainframe
C
Fiber optic cable
11.3.2
Connecting multiple mainframes using the Master/Slave card
The multiple mainframe Master/Slave setup only works in a star configuration.
All Master/Slave fiber optic cables are on one side connected to the Master
mainframe while the other side connects to one of the Slave mainframes.
The master mainframe needs to be extended with one or more optional option
carrier cards (G081). Each option carrier card can be extended with one or two
Master output cards (G083).
GEN3t
I3995-3.1 en HBM: public
163