B.5
Card adapters
B.5.1
Artificial star adapter
G068: Artificial Star Adapter (Option, to be ordered separately)
The artificial star adapter creates an artificial star point to measure 3 phase signals
Maximum input voltage
1000 V DC (707 V RMS) between each of the phases
Inputs
3; 4 mm safety banana plugs
Outputs
6; 4 mm safety banana pins; plugs straight into GN610/GN611/GN610B/GN611B cards
Artificial star N
Reference plug only. Not to be used as input
Safety
Compliant with IEC61010
-
1 600 V RMS CAT II
Application use
The 3 phase signals L1, L2 and L3 can be connected with inputs L1, L2, L3 of the artificial
star adapter. The connection N* is the voltage present on the artificial “star point”.
L1
L1
out
L2
out
L3
out
L2
L3
N*
300k
Ω
300k
Ω
300k
Ω
250pF
250pF
250pF
Figure B.18:
Electrical schematic
Weight
170 g (6 oz)
Material housing
Polyurethane, vacuum resin casting
Setup
Two boxes can be plugged into a single GN610/GN611/GN610B/GN611B card
Two or more GN610/GN611/GN610B/GN611B cards with Artificial star adapters fit next to
each other
Temperature range
Operational temperature
0 °C to +40 °C (+32 °F to +104 °F)
Non
-
operational (storage)
-
25 °C to +70 °C (
-
13 °F to +158 °F)
1-G068-2
Artificial Star
R=300kΩ
L1
L2
L3
IN
L1
Umax L1L2 = L2L3 = Umax L3L 1≤ 1000V DC (707V RMS)
L2
L3
OUT
N*
C=250pF
IEC61010
CAT II 600V
CAT II 600V
1kV DC
L1
L2
L3
N*
19.0 mm
(0.74”)
24.0 mm
(0.94”)
17.0 mm
(0.66”)
16.0 mm
(0.62”)
106.7 mm (4.20”)
101.9 mm (4.01”)
Figure B.19:
Artificial star adapter
GEN3t
I3995-3.1 en HBM: public