7.8
DC power output
POWER
A
Figure 7.16:
DC power output
A
DC power output
The system has a DC power output connector to power additional systems
requiring a DC input power.
The connection is set up to connect a QuantumX system to the GEN series
mainframe directly. It uses a QuantumX compatible connector.
WARNING
QuantumX can connect multiple systems together and attach them all to
one power supply. As the power of the GEN series mainframe is limited,
make sure not to exceed the maximum power output of this connector.
For more information on using the QuantumX together with GEN series
mainframes, please refer to chapter "Synchronizing GEN series and QuantumX
using PTP" on page 876.
DC Power Output
Connector type
ODU, G81L0C
‐
P08LFG0
‐
0000
Mating connector type
ODU, SX1LOC−P08MFG0−0000
Connector pinning
QuantumX compatible; only GND and PWR
signals connected
Output Power
15 Watt
Output Voltage
> 11 V; Typically 11.5 V to 12 V
Maximum Output Current 1.4 A; Limited current and short circuit protected
GEN3t
104
I3995-3.1 en HBM: public