Select the bridge arm to operate switch S5 in Figure 12.31 on page 290:
l
In the sheet, use the
Active gage
column to select between
Positive
(A
-
C) or
Negative
(C
-
D).
l
In the simplified block diagram, click on the
Remote calibration
select
switch (
B
in Figure 12.39 on page 297) to switch between the
two gages.
Shunt verification
-
Procedure
Once all the wiring and resistors have been set up correctly, a shunt verification
can be done.
Shunt verification preparation
1
In Perception, go to the
Settings
sheet.
2
In the task pane, select the
Bridge
in the
Input
section.
3
Select one or more channels.
4
Switch
Excitation ON
.
5
Select an
Excitation voltage.
6
In the task pane, select
Shunt Verification
in the
Sensors
section.
7
Select one or more channels.
8
Select between internal or external shunt usage: In the
Internal shunt
column, enable internal to use an
internal
shunt or clear the option to
select an
external
resistor. Select the appropriate value setting as
described earlier or click on the switch in the diagram (
G
9
Select the bridge arm: Use the
Active gage
column to select between
Positive
(A
-
C) or
Negative
(C
-
D) or click on the switch in the diagram (
I
in
The actual shunt verification is done using the shunt verification dialog.
GEN3t
I3995-3.1 en HBM: public
309