Analog Output GN401 (Receiver)
Channels
4; 1 per transmitter channel (GN110, GN111, GN112 and GN113)
Connector
4; Metal BNC, one BNC per channel on receiver front panel
Conversion
100 MS/s D
-
to
-
A converter per channel
DAC Resolution
14 bit (0.006%)
Outputs
Output filter
Lowpass 10 MHz @ – 3 dB; 6
th
order Bessel reconstruction filter
Output impedance
13 Ω typical
Calibrated Full Scale Output level
±
5 V; 1 MΩ load
Power Requirement GN110 and GN111 (Transmitter)
Battery powered
Maximum 2 removable batteries possible
Note
Use HBM approved batteries only. See option G034 for approved battery details.
Power consumption
6 VA typical, 8 VA maximum
Operation Time (using G034 batteries)
30 hours; 2 batteries installed (15 hours; 1 battery installed)
Perception software can activate a low power sleep mode to extend the operation time
Power Requirement GN112 and GN113 (Transmitter)
Power supply
115/230 V AC @ 47
-
63 Hz (manual voltage selector)
Power consumption
12 VA maximum
Power supply isolation
Protective ground connected
0 V, both sides grounded
Protective ground not connected
1.8 kV RMS (IEC 61010
-
1:2010)
Requires a protected LAB environment and EN50191:2000 compliant work procedures
Fuse(s)
2 x 250 mA; Slow blow
Battery
12 V @ 300 mAh; Internal, rechargeable, NiMH
Battery back
-
up time
5 minutes (with new and fully charged battery)
GEN3t
436
I3995-3.1 en HBM: public