Analog Input
Channels
4
Connector
2 per channel; Metal BNC
Input type
Balanced differential inputs
Input coupling
Single
-
ended positive, single
-
ended negative and differential
Signal input coupling
Coupling modes
AC / DC / GND
AC coupling frequency
1.6 Hz (
±
10%);
-
3 dB
-60
-50
-40
-30
-20
-10
0
0.001
0.01
0.1
1
10
100
M
ag
n
it
u
d
e
[d
B
]
1.6 Hz AC coupling response [dB]
0
10
20
30
40
50
60
70
80
90
100
0.001
0.01
0.1
1
10
100
M
ag
n
it
u
d
e
[%
]
Frequency [Hz]
1.6 Hz AC coupling response [%]
Frequency [Hz]
Figure A.51:
Representative AC coupling response
Impedance
2 * 1 MΩ (
±
2%) // 23 pF (
±
10%)
Ranges
±
20 mV,
±
50 mV,
±
100 mV,
±
200 mV,
±
500 mV,
±
1 V,
±
2 V,
±
5 V,
±
10 V,
±
20 V,
±
50V and
±
100 V
Offset
±
50% in 1000 steps (0.1%)
±
100 V range has fixed 0% offset
DC Offset error
Wideband
0.1% of Full Scale
±
100 μV
Bessel filter
0.1% of Full Scale
±
100 μV
Offset error drift
±
(50 ppm + 20 μV)/°C (
±
(30 ppm + 12 μV)/°F)
DC Gain error
Wideband
0.1% of Full Scale
±
100 μV
Bessel filter
0.1% of Full Scale
±
100 μV
Gain error drift
±
100 ppm/°C (
±
60ppm/°F)
Maximum static error (MSE)
Wideband
0.1% of Full Scale
±
100 μV
Bessel filter
0.1% of Full Scale
±
100 μV
RMS Noise (50 Ω terminated)
Wideband
0.05% of Full Scale
±
100 μV
Bessel filter
0.05% of Full Scale
±
100 μV
Common mode (referred to system ground)
Ranges
Less than or equal to
±
1 V
±
2 V to
±
10 V
Larger than or equal to
±
20 V
Rejection (CMR)
≥ 70 dB
≥ 60 dB
≥ 60 dB
Maximum common mode voltage
4 V DC
40 V DC
250 V DC
Input overload protection
Overvoltage impedance change
The activation of the overvoltage protection system results in a reduced input impedance.
The overvoltage protection is not active for as long as the input voltage remains less than 200% of the
selected input range or 250 V, whichever value is the smallest.
Maximum nondestructive voltage
±
125 V DC; Ranges ≤
±
1 V
±
250 V DC; Ranges >
±
1 V
Overload recovery time
Restored to 0.1% accuracy in less than 40 ns after 200% overload
Restored to 10% accuracy in less than 20 ns after 200% overload
GEN3t
I3995-3.1 en HBM: public
479