It is important to know that sweeps are shown with the trigger aligned at t
0
. The
net effect is that triggers received with a time delay are actually time shifted in
the software, as if these triggers had happened 100% synchronous in time. If
triggers are not 100% synchronous in all systems, signal phase shifts are
introduced by this effect.
Mainframe 1
Mainframe 2
Mainframe N
Trigger origin (t
0
)
Transferred Trigger
Transferred Trigger
t
trigger
Figure 14.5:
Triggering due to trigger transfer delays
The measured signals above appear within Perception as follows:
Mainframe 1
Mainframe 2
Mainframe N
t
0
Figure 14.6:
Signal phase shift introduced by t
0
alignment caused by trigger
transfer delays
GEN3t
372
I3995-3.1 en HBM: public