G906: Passive, Single
-
Ended Voltage Probe (Option, to be ordered separately)
To be used with single
-
ended amplifiers or with differential amplifiers in single
-
ended mode
100 * Uin
Uin
Passive probe
(+)
GND
GND
Single-ended amplifier
900K
10
0K
56
0K
1M
(+)
(+)
(-)
4.5M
49.5M
Figure B.93:
Block diagram passive, single
-
ended voltage probe
Isolation
Not supported
Capacitive compensation range
10 to 50 pF
DC In
-
accuracy
2%
Divide factors
1000:1
Probe impedance (connected to channel)
100 MΩ
-
3 dB Bandwidth
100 MHz
Maximum input voltage
14 kV RMS @ 50/60 Hz, 20 kV DC
Probe cable length
Typically 3 m (9.8 ft)
Probe weight
Typically 465 g (16.4 oz)
Probe operating temperature range
0 °C to +50 °C (32 °F to 122 °F)
Original manufacturer's part number
PMK PHV 4002
-
3
PM
K-14K
V
AC
35 mm
(1.38")
284,4 mm
(11.2")
Figure B.94:
Dimensions
To be used with single
-
ended amplifiers or with differential amplifiers in single
-
ended mode
Figure B.95:
Probe and probe accessories
Note The compensation range of the G906 probe does not match that of GN1610/GN1611/GN3210/GN3211 cards.
For the limited bandwidth of GN1610/GN1611/GN3210/GN3211, this has no noticeable effects. When using
the G906 probe in combination with this card select the sensor "G906_NoCapacitiveCheck" from the Perception
Sensor Database.
GEN3t
838
I3995-3.1 en HBM: public