2.) Unscrew four screws (3-6) on the right side and (7-10) on the left to remove the speakers. Refer to
3.) Unscrew four screws (7-10) to remove the speaker on the right.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-3-5-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.
Figure 8-14 Speaker Assy Disassembly
Table 8-6
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-3-5. Equipment
passed all required tests and is ready for use. “