Chapter 8 Replacement Procedures
8 - 17
4.) Disconnect three connectors (6-8) from the HUB board. Refer to Figure 8-19.
5.) Disconnect a connectors (9). Refer to Figure 8-20.
6.) Carefully lift the Keyboard assy from the system.
Figure 8-19 Disconnect 3 connectors from the HUB
Figure 8-20 Disconnect two connectors
H U B A s s y
6
7
8
H U B A s s y
6
7
8
9
9