Chapter 8 Replacement Procedures
8 - 105
8-8-1-5 BEP Cable Connection with Console( In case of change BEP 1 into BEP2 or BEP3)
8.) Connect CON21 Connector to BEP.
9.) Connection COM1 Connector to Serial Port on BEP.
NOTE:
COM1 Cable Connect to Motherboard Serial Port ( D-SUB 9 Pin Male )
Figure 8-97 CON21 cable connection
Figure 8-98 COM1 cable connection
C O N 2 1 c a b le
C O N 2 1 c a b le
COM1 cable