GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
4 - 16
Section 4-3 - General Procedure
4-3-11
Backup and Restore Database, Preset Configurations and Images
4-3-11-1
Formatting Media
1.) To format the backup media, CD-ROM or MOD, select the UTILITY button on the Menu panel.
2.) Select CONNECTIVITY, then TOOL( in case of BT05 SW, REMOVABLE MEDIA )on the
monitor screen. Properly label and Insert the backup media.
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure 4-31. It is best to use all capital letters with no
spaces or punctuation marks.
5.) The Ultrasound system displays a pop-up menu, as shown in Figure 4-32. When the formatting
has been completed, press OK to continue.
6.) If desired, verify that the format was successful by returning to
Utility>Connectivity>Tools (in
case of BT05 SW, Removable Media)
and selecting VERIFY as shown in Figure 4-31.
NOTICE
Avoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator
unlocked and the MOD susceptible to damage if moved. If forced to use this method, reboot the
system, then insert and eject a known good disk using one of the other manual ejection
methods.
Figure 4-31 Format and Verify Media
Figure 4-32 Format Successful Pop-up Menu
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...