Section 5-4 - Front End
5-4-1
RLY
RLY ASSY contains of 3 phased array probe connectors and provides switchable connection between
probes and 64ch transmitters/receivers.
NOTE:
There is no numeric order in the probe port, system recognizes the first probe slotted on the
port as the probe #1, regardless of the probe port location.
This board also has interface for a single CWD probe connector.
The main function of RLY ASSY is as follows.
-
3-to-1 selectors for three probes.
-
Interface with JUSC bus (control bus)
-
Interface with FEC ASSY for IIC bus.
-
Interface with DCWD Connector.
-
Supply/Cut control and failure detection of supply voltage for Mux circuit in a Probe.
-
Device: Mechanical Relay
-
Switch: 128ch 3:1 selector
-
Temperature sensor in this assy detects temperature of RLY assy.
Figure 5-5 RLY Block Diagram
Probe A
Probe C
DCWD
Probe I/F
RELAY
128
128
Probe B
128
128
RELAY
128
Probe ID
I/F
Control
Circuit
Control
FPGA
MUX
64
SHV
To LMT
Temperature
Sensor
JUSC I/F