GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 197
8-11-3-2
Setting Computer Name
NOTE:
If the "Save" is NOT performed, you can NOT enter Patient screen and NOT perform Export/
Import
1.) Press Utility-> Connectivity
2.) Operator Login Window appears. Select ADM(Administrator) for operator field.Click OK button
3.) Click on Tcpip tab
4.) Check if Computer Name, IP Address, Subnet Mask, and other parameters are proper
5.) Click on Save button
6.) Click OK for confirmation dialog box
7.) The system will power OFF automatically.
8.) Turn on the system for functional checks
Figure 8-237 Set Computer Name
Figure 8-238 TCPIP tab
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...