GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 38
Section 8-4 - Keyboard Block
2.) Disconnect two connectors and unscrew 8 screws. Refer to Figure 8-40.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-11-5
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-40 Unscrew 8 screws
Table 8-18 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
Equipment passed all required tests and is
ready for use. “
Section 4-11-10
Menu Key Assy
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8