GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 147
8-8-3-4 Removal Procedure
(cont’d)
15.)Check if CD R/W is properly installed in the Maintenance Mode. (Check the CR R/W is recognized
by the system in the MS windows navigator)
16.)Press and hold the Power Button for 20 sec. to execute manual system shut down.
17.)Enter BIOS Set up mode again.
18.)Reset the Diskette Controller to
[Disable]
in the Advanced Menu screen.
19.)Press "F10 - SAVE AND EXIT" to save the BIOS set up.
20.)Perform the following functional tests. If all are successful, include the debrief script provided below.
8-8-3-5
Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-48 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-8-3. Equipment
passed all required tests and is ready for use. “
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...