LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 154
Section 8-9 - Power Block
3.) Unscrew ten (10) screws (11-20) to remove AC Power assy. Refer to Figure 8-173.
4.) Remove the AC Power Assy.
Figure 8-173 Removing the AC Power Assy
WARNING
WARNING
The weight of the AC Power Assy is approximately 20 kg. Two person is needed in the
next step.
11
12
13
15
16
17
14
18