GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
4 - 34
Section 4-11 - Keyboard Function Checks
4-11-9
OP Panel LCD Encoder Assy Validation
1.) Boot up the system and run the scfpapitest program.
2.) Press and Rotary LCD Encoder.
3.) Check the log dialog box in scfpapitest program.
NOTICE
Follow up the procedure from step 1 to step 7 of Section 4-11-1 to run the scfpapitest program.
Button or Rotary Name
The Name displayed in Log
1 Button
PushEncoder 1 of Touchpanel Button
2 Button
PushEncoder 2 of Touchpanel Button
3 Button
PushEncoder 3 of Touchpanel Button
4 Button
PushEncoder 4 of Touchpanel Button
5 Button
PushEncoder 5 of Touchpanel Button
1 Rotary
PushEncoder 1 of Touchpanel Rotary
2 Rotary
PushEncoder 2 of Touchpanel Rotary
3 Rotary
PushEncoder 3 of Touchpanel Rotary
4 Rotary
PushEncoder 4 of Touchpanel Rotary
5 Rotary
PushEncoder 5 of Touchpanel Rotary
Brightness Rotary
Brightness PushEncoder Rotary
Table 4-26 Button and Rotary Table
NOTICE
The Name displayed in log while rotating the rotary button, The menu Encoder is working correctly.
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...