6 - 2
Section 6-3 - Power Supply Adjustment
6-3-1-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-2-2 on
page 5.
•
Remove the AC Power assy from the System
6-3-1-4
Removal Procedure
1.) Unscrew four (4) screws (1-4) to remove Left side cover of AC Power assy.
2.) Disconnect three (3) connectors from the rear side of the LV Unit. Refer to.
3.) Unscrew four (4) screws (8-11) to remove Right side cover of AC Power assy.
4.) Unscrew four 4 screws and disconnect twenty (20) connectors from the front side of the LV unit
using long-nose gripper.
5.) Remove the LV unit.
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
Figure 6-1 Disconnect 3 connectors
Figure 6-2 Disconnect 20 connectors
Connector
Connector
Connector
Screw
20 Connectors
Screw
Screw
Screw