Chapter 2 Pre-Installation
2-5
2-2-4
Scan Probe Environmental Requirements
Operation:10° to 35° C
Storage:--10° to 60° C
NOTE:
Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
Replace all screws, RF
gaskets, covers, cores
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end. Install the
shield over the front of card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interface with the ultrasound signals.
Replace broken RF gaskets
If more than 20% or a pair of fingers on the RF gaskets are broken, replace the gaskets.
Do not turn on the unit until any loose metallic part is removed.
Do not place labels where RF
gaskets touch metal
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. Or, if a label has been found in such a position, move the label.
Use GE specified harnesses
and peripherals
The interconnect cables are grounded and require ferrite beads and other shielding. Also,
cable length, material, and routing are all important; do not change from what is specified.
Take care with cellular phones Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Properly dress peripheral
cables
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor
cables to the frame.
NOTICE
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES
OF -20 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
Table 2-5
EMI Prevention/abatement
EMI Rule
Details