GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 170
Section 8-10 - Software Loading Procedure
8-10-5-3
Saving Connectivity
Before starting upgrade, write down the
Connectivity Setting
for back-up.
1.) Press
Utility > Connectivity
and click the
Tcpip
tab.
2.) Write down the following parameters:
Parameters
Descriptions (Entry for Stand-Alone System)
Computer Name
Enable DHCP
IP-Address
Subnet Mask
Default Gateway
Remote Archive IP-Addr
Remote Archive Name
Note : For a stand-alone system, preset values of IP address, subnet mask, and default gateway, shown
above, must entered.
Figure 8-189 Connectivity
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...