Chapter 6 Service Adjustment
6 - 3
6-3-2
Adjustments Procedures
1.) Using the following VRs, DC output can be adjusted.
2.) Each DC voltage should meet the voltage range as table. if output voltage is out of range, adjust
the output level with the VR on each module.
3.) Clockwise direction of VR increase voltage level, counter clockwise of VE decrease voltage output
Table 6-33 DC Output Specification for LV Unit
VR
Voltage Range Spec on each Connector
VR1
6.17 ~ 6.21V
VR2
3.43 ~ 3.47V
VR3
6.23 ~ 6.27V
VR4
-6.07 ~ -6.09V
VR5
-15.00 ~ -15.04V
VR6
5.19 ~ 5.23V
VR7
15.00 ~ 15.04V