Section 8-6 - Body Block
Section 8-6
Body Block
8-6-1
Rear Handle (FRU No. 400)
Purpose: This is a description on how to remove and replace the Rear Handle.
8-6-1-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-6-1-2
Needed Manpower
•
1person, 25 m travel
8-6-1-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-6-1-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Rear Cover (FRU No.302). Refer to section 8-5-3 on page 45.
4.) Remove the Top Cover (FRU No.304). Refer to section 8-5-5 on page 49.
5.) Unscrew four (4) screws (1-4) to remove peri rear bracket. Refer to Figure 8-65.
Figure 8-65 Removing the Peri Rear Bracket