GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 81
8-6-3
Front Caster (FRU No. 403)
Purpose: This is a description on how to remove and replace the Front Caster.
8-6-3-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
•
Lifter (180Kg)
8-6-3-2
Needed Manpower
•
1 persons, 15 m travel
8-6-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Lift the system before proceed.
8-6-3-4
Removal Procedure
1.) Unscrew eight (8) hexagonal bolts (1-8).
2.) Remove the Front Caster. Refer to Figure 8-75.
Figure 8-75 Removing the Front Caster
1
2
6
8
7
5
4
3