Section 8-5 - Covers
8-5-12
OP Curtain Cover (FRU No. 311)
Purpose: This is a description on how to remove and replace the OP Curtain Cover.
8-5-12-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-5-12-2
Needed Manpower
•
1 person, 25 m travel
8-5-12-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-5-12-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Front Base Cover (FRU No.305). Refer to 1.) on page 8-40.
4.) Remove the Front Cover (FRU No.306). Refer to section 8-5-7 on page 53.
5.) Remove the OP Side L Cover (FRU No.307). Refer to section 8-5-8 on page 54.
6.) Remove the OP Side R Cover (FRU No.308). Refer to section 8-5-9 on page 56.
7.) Remove the OP Bottom Cover (FRU No.310). Refer to section 8-5-11 on page 60.
8.) Unscrew 8 screws from the OP Curtain Cover. Refer to Figure 8-59.
Figure 8-59 Unscrew 6 screws to remove the OP Curtain Cover
OP Bottom Cover
OP Curtain Cover
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Summary of Contents for LOGIQ 5 PRO
Page 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Page 3: ......
Page 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Page 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...