Chapter 8 Replacement Procedures
8 - 87
8-6-6
Air Filter Set (FRU No. 406)
Purpose: This is a description on how to remove and replace the Air Filter Set.
8-6-6-1
Tools
•
Not Required.
8-6-6-2
Needed Manpower
•
1 person, 5 m travel
8-6-6-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-6-6-4
Removal Procedure
1.) Remove Air Filter as illustrated below.Refer to Figure 8-78.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-6-6-5
Mounting procedure
Install the new parts in the reverse order of removal.
Figure 8-78 Removing the Air Filters
Table 8-41 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
2380207
, Rev 7+, Section 8-6-6. Equipment
passed all required tests and is ready for use. “