Chapter 8 Replacement Procedures
8 - 49
8-5-5
Top Cover (FRU No. 304)
Purpose: This is a description on how to remove and replace the Top Cover.
8-5-5-1
Tools
•
Common pillips screwdrivers
8-5-5-2
Needed Manpower
•
1 person, 20 m travel
8-5-5-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-5-5-4
Removal Procedure
1.) Remove the Left Cover (FRU No.300). Refer to section 8-5-1 on page 41.
2.) Remove the Right Cover (FRU No.301). Refer to section 8-5-2 on page 43.
3.) Remove the Rear Cover (FRU No.302). Refer to section 8-5-3 on page 45.
4.) Unscrew two screws (1-2) from the top side of the Top cover. Refer to Figure 8-47.
Figure 8-47 Unscrew 2 screws from the Top cover
(1)
(2)
Summary of Contents for LOGIQ 5 PRO