4 - 4
Maintenance
K0260 Issue No. 8
4.4
Routine Maintenance
WARNING: S
WITCH
OFF
AND
DISCONNECT
THE
POWER
SUPPLY
BEFORE
STARTING
ANY
MAINTENANCE
TASK
.
Carrying out the maintenance tasks detailed 4.3.
Servicing Procedures
The following procedures provide instructions to test and replace items for the operator. Return the
unit to the repair depot for further testing and replacement of items.
Replacing the output connector o-ring
After inspection as detailed in maintenance task B, carry out the following if the o-ring
is worn or damaged:
Carefully remove the o-ring from the
small groove at the top of the
connector.
Fit a new o-ring in the small groove at
the top of the connector.
Make sure the o-ring is tight in the
groove and not damaged after fitting.
Note:
Damage to this o-ring causes leaks.