2 - 6
Installation
K0260 Issue No. 8
2.4
Electrical Connection
WARNING: VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50 VOLTS DC, IN CERTAIN
CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE TAKEN WHEN WORKING ON LIVE,
EXPOSED CONDUCTORS.
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
Open end cables
Only a qualified technician (see page i) must fit a connector to a cable supplied by GE (see Table 2-1,
item 3).
C
AUTIONS
:
1
T
HE
SUPPLY
MUST
PROVIDE
CONNECTION
TO
A
PROTECTIVE
GROUND
TERMINAL
. T
HE
UNIT
MUST
,
AT
ALL
TIMES
,
BE
CONNECTED
TO
THE
SUPPLY
EARTH
(
GROUND
).
2
T
HE
POWER
SUPPLY
CABLE
AND
CONNECTOR
MUST
BE
CORRECTLY
RATED
FOR
THE
POWER
SUPPLY
.
Note:
The ADTS 505 is normally supplied with an approved power supply cable for use in the country
of delivery. This can limit the maximum supply voltage that can be safely used.
e.g. a NEMA 5-15P terminated cable, for use in the U.S.A., is approved for a maximum of 125 V
ac; it must be replaced for a higher supply voltage.
Make sure that the power supply is off before connecting the power cable.
If required, connect the hand terminal to the connector on the front panel.
Note:
Connecting the hand terminal to the unit disables the front panel key-pad.
Fuses
The two fuses, located in the holders and mounted on the front panel, protect the unit. The
fuses are connected in the live and neutral supply circuit and are T5.0A/250V HBC type.
Note:
Only replace the fuses with the correct type, see Figure 2-1 item1, to replace see section 4-5.
External earth/ground connection
An external earth (ground) cable may be connected to the stud on the front panel of the unit
providing integrity of the earth (ground) connection.
F
IGURE
2-2 P
OWER
SUPPLY
LABEL
Pin
(ADTS)
European
Colour
U.S.
Color
Function
1
Brown
Black Live
4
Blue
White
Neutral
Centre
Green/Yellow
Green
Protective Earth (Ground)
100 / 120 / 230V ~ 50 / 60Hz 200VA
115V ~ 400Hz 200VA
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...