5 - 14
Testing and Fault Finding
K0260 Issue No. 8
5.6
Fault Finding
System Messages
In the event of a malfunction, the built-in, self-test and diagnostic system displays a message
and a code. The message heading
Error
indicates a fault or condition that interrupts normal
operation. The following tables list the error messages with the probable cause and action to
be taken.
If the display shows a message not listed in the tables the unit must be switched off. If the
display shows the same non-operational message after switching on, the unit should be
returned to the repair depot.
T
ABLE
5-3 E
RROR
M
ESSAGES
Note:
In No. 56, this safety feature opens the zero valve. If this happens constantly, check for leaks.
No.
Message
Probable Cause
Action
7
Autoleak recovery
activated
Excessive leak rate detected in Ps/Qc
pressures
Check hoses and aircraft systems for
leaks
31
Test set must be at ground Invalid user request
Select Go to Ground first
32
Pressure leak in Ps.
Leaking Ps channel pressure control
valve.
Return unit to repair depot.
33
Pressure leak in Pt.
Leaking Pt channel pressure control
valve.
Return unit to repair depot.
36
Ps Aim Overshoot.
Connected system leak rate
changing.
Check hoses and aircraft systems for
leaks.
37
"Ps aim value
unobtainable".
Leak rate on Ps too high to control.
Check port caps, hoses and
connectors for leaks.
40
Pt Aim Overshoot.
Connected system leak rate
changing.
Check hoses and aircraft systems for
leaks.
41
"Pt aim value
unobtainable".
Leak rate on Pt too high to control.
Check port caps, hoses and
connectors for leaks.
4
Vacuum leak in Ps.
Leaking Ps channel vacuum control
valve.
Return unit to repair depot.
4
Vacuum leak in Pt.
Leaking Pt channel vacuum control
valve.
Return unit to repair depot.
4
Ps pressure control valve
cannot be calibrated.
Valve fault.
Return unit to repair depot.
56
Stabilising Qc to Zero.
Possible negative airspeed detected,
possible sudden leak.
System zero valve opened to stop
negative airspeed condition.
57
Current Pressure Limits
Exceeded.
Limits for Ps or Qc exceeded.
Make sure pressures are within limits,
check leak rates.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...