3 - 14
Operation
K0260 Issue No. 8
To select a limit set (User)
Press SETUP, the display shows Setup, [Limits]
[xxxxxx], where xxxxxx is the current name of
the selected limit set. Select [Limits] [xxxx] F3.
The display highlights the current set of limits
in use by the name in the list. The current limit
values can be inspected by pressing [View
Details] F4. The display shows the current
limit values and divides these values into
aeronautical (Page #1) and pressure (Page
#2).
The ARINC 565 limit can be selected ON or
OFF in a set of user defined limits, When
selected, the values entered and/or achieved
in control mode are checked within the
operating envelope of the ARINC 565 limits,
see section 6.
To select another set of limits from the list
press [Select Limits], F5. Use [ ] or [ ] to
move through the list.
To edit a limit set (Supervisor)
To edit or make a new set of limits press [Enter
PIN], F3 and enter the PIN. Use [ ] or [ ] to
move through the list. Each set of limits is
identified by it's name. New sets of limits can
be created and existing sets of limits can be
renamed or values changed. Refer to section
6 for more details. The user-named sets, can
be edited by pressing [Select Limit [xxxxxx],
then Modify Details or Modify Name.
Select [Save Settings] to store the new details
or name to be used on the next power-up of
the system.
Notes:
1
For temporary changes press QUIT/CLEAR
in place of [Save Settings].
2
The factory-defined limit sets, “STANDARD“,
"MAX" and "CIVIL" cannot be changed.
Limits
t
s
View
Details
Select
Limits
Enter
PIN
ALT
Ps
SPEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEASURE
CONTROL
HELP
SETUP
USER 1
USER 2
USER 3
USER 4
USER 5
MAX
STANDARD
CIVIL
Active Limit Set: CIVIL
Available Preset Limits
Available User Limits
t
s
Select
Limit
Limits
Save
Settings
Modify
Details
Modify
Name
User
Supervisor
ALT
Ps
SPEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEASURE
CONTROL
HELP
SETUP
Min ALT
-1000 ft
Min CAS
Max Mach
1.0 Mach
Max ROC
6,000 ft/min
Max ALT
50,000 ft
0.0 knots
Max CAS
450 knots
Alt Cor.
0 ft
ARINC Limits
OFF
Page #1
Details
Next
Page
Next
Page
User
Supervisor
t
s
Details
Modify
Units
more
(1 of 2)
ft kts
ft/min
ALT
Ps
SPEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEASURE
CONTROL
HELP
SETUP
Rt
Min Ps
0.0 mbar
Max Ps
1050.406 mbar
Min Qc
0.0 mbar
Max Qc
368.01 mbar
Ps
105.00 mbar/min
Rt Qc
90.0 mbar/min
Page #2
Details
Next
Page
Limits
more
(1 of 2)
m km/h
m/min
mbar
mbar/min
User
Supervisor
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...