![background image](http://html.mh-extra.com/html/ge/druck-adts-505/druck-adts-505_user-manual_132652031.webp)
Druck ADTS 505 User Manual
3 - 3
K0260 Issue No. 8
3
3.3
Quick Reference
Key/selection
Function and comments
F
IGURE
3-1 K
EY
-
PAD
AND
D
ISPLAY
KEY-PAD FUNCTION
the display shows the main pressure display (Leak Measure or Control mode) with
normal operation key functions.
F1-F6
Function keys for menus
ALT/Ps
Altitude (Aeronautical units) or Ps (pressure units)
SPEED Qc
Airspeed (Aeronautical units) or Qc (pressure units)
EPR
Engine Pressure Ratio (pressure units only)
ROC/RATE Ps
Rate of Climb (Aeronautical units) or Rate of Ps (pressure units)
LEAK MEASURE/CONTROL
Switches between measure mode (for leak testing) and control mode
RATE
Rate of change of Pitot or Mach parameter, press parameter then RATE (read only)
GROUND
Controls Ps to atmospheric pressure and Qc to zero both at current rates of change
SETUP
Changes functions, limits and units, if [save/lock] pressed changes are permanent
HELP
Press HELP for further information on current selection/display
0-9
Number entry
-/000
Minus sign for first number entry 000 (thousand) if not first number of entry
CLEAR/QUIT
Clear number entry - quit from menu, HELP screen or clear warning message
ENTER
Complete number entry
DELETE
Removes the last number or character entered
CLEAR/QUIT + DELETE =
ABORT, all operations stop, the ADTS 505 restarts from a normal power-up
sequence, including safe equalising of pressures between the test set and the
connected system
100 / 120 / 230V ~ 50 / 60Hz 200VA
115V ~ 400Hz 200VA
T5.0A
250V
HBC
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...