Fault
Symptom
Probable cause
Action
F1
Power supply indicator
does not light.
Faulty power supply outlet.
Unserviceable power supply front
panel fuse.
Faulty power supply cable.
Internal power supply fault.
Check power supply.
Check/replace fuse.
Check/replace fuse.
Replace cable and re-test.
Record fault and return ADTS 505
to depot for repair.
F2
Power supply indicator
lights, no display.
Internal fault.
Record fault and return ADTS 505
to depot for repair.
F3
No self-test message.
Internal fault.
Record fault and return ADTS 505
to depot for repair.
F4
Display shows error
message.
Error detected.
Refer to fault finding tables.
F5
Display does not show
Leak Measure mode.
If display stops at Thermal Test.
System component fault.
Allow ADTS 505 to reach ambient
temperature.
Record fault and return ADTS 505
to depot for repair.
F6
Fails SST.
Internal leak.
Internal fault.
Refer to further testing 5.5.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...