Druck ADTS 505 User Manual
3 - 11
K0260 Issue No. 8
3
3.5.3 Control or Measure Parameter
To change the displayed Measure parameters:
Press
SETUP
[Displays], [Single, Dual, Quad], press
CLEAR/QUIT
repeatedly until the display goes
to Leak Measure Mode.
Measure parameters:
Press the parameter key
e.g., press
SPEED/Qc
to display airspeed.
To change the displayed rate parameter:
Press the associated parameter key followed by the rate key.
e.g., display airspeed rate, press
SPEED/Qc
then
RATE
.
ROC/RATE Ps
may be directly pressed without first pressing
ALT/Ps
.
In Leak Measure mode, pressing the parameter and rate keys changes the order that the display
shows the parameters.
Control Aim
A new control aim can be entered using the numeric keys. Each digit is displayed as it is pressed.
The existing aim is replaced when the first digit of the new aim is pressed. If an error is made during
the entry of data, press
CLEAR/QUIT
to restore the original aim.
Press
ENTER
to action the new aim.
Note 1:
The
-/000
key can be used as a quick way of entering thousands. If this key is pressed first (i.e.
before a number is entered) the value becomes negative, press again the value becomes positive.
Note 2:
The ADTS 505 must be in control mode to enter a new aim (current aim displayed). If the aim
field shows "Leak Measure" press
LEAK MEASURE/CONTROL
to enter control mode.
3.5.4 Leak Testing the ADTS 505
It is important to check that the ADTS 505 and the connecting hoses do not leak. Before use, a leak
check should be carried out as follows:
Connect the Pitot and static hoses (to front panel of ADTS 505). Temporarily seal the free ends of
the hoses.
Set the display to quad, see 3.4.1. Set the units to feet and knots, see 3.5.5.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...