3.7.5 Leak Testing
General
Apply pressure to the system under test as detailed in Testing the Aircraft Static or Pitot system.
Before starting the leak test, check that the ADTS 505 is operating within range of the set limits.
Example
To leak test a pitot system to 300 knots at a maximum rate of 300 knots/min and a static system
to 30,000 ft at a maximum rate 5,000 ft/min. Tests with a wait time of 5 minutes and test time of
1 minute. Maximum pitot system leak rate allowed 5 kts/min or less. Maximum static system leak
rate 200 ft/min or less.
Procedure
1.
Select F2 [Units], from the Main Menu, select F1 [ft, kts, kts/min]. Press
CLEAR/QUIT
to return to the
main menu.
2.
Press
SETUP
, select F1 [Display], press F3 [Quad] and then press
ALT/Ps
,
ROC/RATE Ps
,
SPEED/Qc
and
RATE
so that the display shows these parameters in the most appropriate order.
3.
Press F3, [Rate Timer], then F3 [Set Wait 5.00], press
ENTER
and F4 [Set Time 1.00], press
ENTER
.
Press F6 to return to the Main menu.
4.
Press the
LEAK MEASURE/CONTROL
to enter control mode. Press
ROC/RATE Ps
and set an Aim of
5000 ft/min. Press
ALT/Ps
and set an Aim value of 30,000 ft and press
ENTER
. The system starts
to control to the new set-point.
5.
Press the
LEAK MEASURE/CONTROL
to enter measure mode. Press F1 [Rate Timer] and press F1
[Start Timing].
6.
At the end of the timing check the rate of change (leak rate) this should be within the required limits
stated in the Aircraft Maintenance Manual; in this example 200 ft/min or less.
7.
Press the
LEAK MEASURE/CONTROL
to enter control mode. The preset rate of speed is set at 300
knots/min. Press
SPEED/Qc
and set an Aim of 300 knots. The system starts to control to the new
set-point.
8.
Press the
LEAK MEASURE/CONTROL
to enter measure mode. Press F1 [Rate Timer] and press F1
[Start Timing].
9.
At the end of the timing check the rate of change (leak rate) this should be within the required limits
stated in the Aircraft Maintenance Manual; in this example 5 kts/min or less.
Note:
If further testing is required, apply another pressure to the system under test and repeat the
steps above.
On completion of testing, press the
LEAK MEASURE/CONTROL
to enter control mode, select
GROUND
and wait for the `Safe at Ground' message to be displayed, see 3.7.8., refit the pitot and
static blanking caps.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...