3 - 22
Operation
K0260 Issue No. 8
3.7.7 Engine Pressure Ratio
The ADTS 505 may be used to check EPR sensors and indicators. Use Ps (static) for INLET pressure
and Pt (Pitot) for OUTLET pressure.
To carry out an EPR check, the display must be showing units of pressure measurement e.g., mbar,
inHg, psi.
Note:
A
ltitude correction applies to EPR values make sure this correction does not adversely affect
accuracy. The difference in height between the ADTS 505 and the EPR sensor may be different
to the altitude correction previously set for the aircraft instruments.
To change units:
•
Press F2 [Units].
•
Select required pressure units, the pressure units are selected in set-up.
•
Use
CLEAR/QUIT
to return to pressure display.
To enter an EPR of 1.8 with inlet pressure of 500 mbar (15 inHg), proceed as follows:
•
If the display shows "Leak Measure", press
LEAK MEASURE/CONTROL
to go to control.
•
Press [Rate Inlet], F2 and set 200mbar/min.
•
Press [Rate EPR], F4 and set 0.8 EPR/min.
•
Press [Inlet], F1 set an aim value of 500 mbar (15 inHg).
•
Select [EPR] F3, set an aim value of 1.8.
•
Wait for the controller to achieve the EPR aim value. Press [Outlet], F5 to confirm 900
mbar.
Note:
EPR testing can also be performed by specifying the actual INLET and OUTLET values. In the
above example, with inlet at 500 mbar, a ratio of 1.8, the outlet value should be 500 mbar x 1.8
= 900 mbar (Pt) [15 inHg x 1.8 = 27 inHg (Pt)].
Some EPR procedures (earlier ECU) may require measurement in aeronautical units:
•
Set to aeronautical units, press
LEAK MEASURE/CONTROL
and set the inlet pressure
(ALT control aim) to the required value.
•
Change units to pressure, press EPR, set the required ratio.
•
Allow the system to achieve the control aim. Exit EPR and return to leak measure
mode. Change units to aeronautical and record the resulting value in knots.
On completion of EPR testing, select
GROUND
and wait for the `Safe at Ground' message to be
displayed, see 3.7.8., refit the pitot and static blanking caps.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...