5 - 12
Testing and Fault Finding
K0260 Issue No. 8
Vacuum leak check
This procedure verifies that the unit is leak tight under vacuum conditions.
(1)
Press
LEAK MEASURE/CONTROL
to enter control mode.
(2)
Enter an ALT/Ps Aim of 100 mbar.
(3)
Enter a Speed/Qc Aim of 0 mbar.
(4)
Wait the required time for temperature stabilization e.g., one minute. Press
LEAK
MEASURE/CONTROL
return to Leak Measure mode.
(5)
Select [Rate Timer] and select [Set Wait 05:00], [Set Time 01:00] and then select [Start
Timing].
(6)
At the end of the timing period, press
ROC/RtPs
to display the measured rate of change
of ALT/Ps.
(7)
Check that the Ps, Qc and Pt rates are less than or equal to ±0.5 mbar/min. If the leak
rate is not achieved, allow further thermal stabilization time and re-test by selecting
[Rate Timer] and select [Set Wait 05:00], [Set Time 01:00].
(8)
Press
CLEAR/QUIT
to exit rate timer display.
Range check
Control the Ps and Qc pressures to the normal limits of operation and make sure that these
are achieved. Use minimum Ps and zero Qc and then maximum Ps and maximum Qc.
Note:
Do not exceed the maximum Pt as shown on the front panel of the unit. Use
maximum available rates of change to minimise test times.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...