5 - 2
Testing and Fault Finding
K0260 Issue No. 8
5.2
Standard Serviceability Test
The following procedure shows if the unit is serviceable and checks functions and facilities of
the ADTS 505. In this procedure:
•
All key presses are highlighted in
BOLD
and shown as identified on the front panel.
•
Key presses inside brackets e.g., [Units], are soft key presses (i.e., function key
selections {F1 to F6} indicated on the screen).
Procedure
Connection and power-on checks.
•
Connect power to the unit.
•
Make sure the blanking caps are fitted to the
Ps and Pt front panel outputs.
•
Set the power supply switch to ON.
•
Check power indicator is on.
•
Check the display shows the first power-up
message.
Check the display all the power-up messages.
•
The ADTS 505 displays any detected errors,
refer to fault finding.
•
Check that the display then changes to show
measured aeronautical or pressure values. This
depends on the display set-up.
N
OTES
1
The displayed values change as atmospheric pressure changes at power-up.
2
The unit requires
15 minutes to
warm-up to
achieve full
accuracy and
stability.
ALT
Ps
SPEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEASURE
CONTROL
HELP
SETUP
SYSTEM STARTING
PLEASE WAIT
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
ADT
S
505
Software Version 1.06
LAST CALIBRATED [DD/MM/YYYY]
THERMAL TEST PLEASE WAIT
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
Software Version 1.06
LAST CALIBRATED [DD/MM/YYYY]
CHECKING FOR VACUUM LEAKS
ADT
S
505
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
Software Version 1.06
LAST CALIBRATED [DD/MM/YYYY]
CHECKING FOR PRESSURE LEAKS
ADT
S
505
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
Software Version 1.06
LAST CALIBRATED [DD/MM/YYYY]
PRESSURIZING PUMPS
ADT
S
505
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
KEEP SIDE AIR VENTS CLEAR
OPEN VALVES
ADT
S
505
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
KEEP SIDE AIR VENTS CLEAR
GROUND SENSORS
ADT
S
505
ALT
Ps
S
PEED
Qc
MACH
Pt
ROC
RATE Ps
GROUND
RATE
7
F6
F5
F4
F3
F2
F1
4
1
_
000
8
5
2
0
9
6
3
.
CLEAR
QUIT
ABORT
DELETE
ENTER
LEAK
MEA
S
URE
CONTROL
HELP
S
ETUP
KEEP SIDE AIR VENTS CLEAR
FINDING VALVE BIAS 1 OF 4 . . .
ADT
S
505
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...