Installation
K0260 Issue No. 8
2.6
Positioning of the ADTS 505
WARNING: IN AN ENCLOSED AREA WITH FUEL VAPOUR PRESENT THIS EQUIPMENT MUST BE
PLACED AT LEAST 0.5 METRES ABOVE FLOOR LEVEL. THIS EQUIPMENT CONTAINS A
D.C. MOTOR WITH BRUSHES THAT COULD CAUSE A SPARK.
C
AUTION
:
T
O
OPERATE
,
PLACE
THE
UNIT
ON
A
HORIZONTAL
SURFACE
(
FRONT
PANEL
UPPERMOST
)
OR
VERTICALLY
(
POWER
SUPPLY
CONNECTOR
UPPERMOST
)
THIS
ALLOWS
THE
WATER
IN
THE
WATER
FILTER
TO
VENT
.
W
ATER
COULD
CONTAMINATE
THE
CONTROLLER
MANIFOLD
AND
AFFECT
CONTROLLER
PERFORMANCE
.
Note:
In control mode, the water drain, located at one end of the unit near the carrying handle, produces
a flow of air and some water. The amount of water depends on the humidity and the operating
time in control mode.
C
AUTION
:
B
EFORE
USE
,
CHECK
THE
WATER
DRAIN
PIPE
,
IT
MUST
BE
FREE
OF
OBSTRUCTION
. W
HEN
IN
CONTROL
MODE
CHECK
THAT
A
SMALL
FLOW
OF
AIR
COMES
OUT
OF
THE
DRAIN
PIPE
.
Notes:
1
When checking for a small flow of air do not block the drain pipe completely this causes a back-
pressure in the pipe and controller instability.
2
If no air flows from the drain pipe, when in control mode, switch off and start again. If no air
flows after a re-start, switch off and do not use the ADTS 505, return the unit to the repair depot.
C
AUTION
:
T
HE
SIDE
VENTS
MUST
NOT
BE
OBSTRUCTED
,
THIS
UNIT
REQUIRES
AN
AIRFLOW
FOR
THE
INTERNAL
COOLING
FANS
.
It is important that the position of the ADTS 505 in relation to the aircraft altitude sensors is
known. An altitude correction must be made to allow for the difference in height between the
reference level and the aircraft's altitude sensors (Figures 2-2 and 2-3). The Reference section
contains details of altitude correction.
WARNING: OBSERVE THE APPROPRIATE SAFETY INSTRUCTIONS AND TESTING PROCEDURES
DETAILED IN THE AIRCRAFT MAINTENANCE MANUALS AND COMPONENT
MAINTENANCE MANUALS.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...