Druck ADTS 505 User Manual
3 - 13
K0260 Issue No. 8
3
3.5.5 Changing the Units of Measurement
The units of measurement can be changed
to units of pressure measurement or
aeronautical units.
To change the units:
•
Select, from the main menu, [Units].
•
Select from the Units Menu:
•
Select [Save Settings] to store as the
units used at the next power up.
•
Press
CLEAR/QUIT
to return to the
main pressure display.
•
In set-up, more units of pressure
measurement can be selected for
each parameter.
3.5.6 Setting Limits
All data entered is checked against minimum and maximum values set for the particular limit set
in use. If these limits are exceeded, the display shows a warning message "Changed aim to current
limit". Press
CLEAR/QUIT
the system automatically modifies the aim values so that the limits are
not exceeded. All limits of associated parameters are checked, if these limits are exceeded the
display shows the warning message and the parameter modified. Similarly, if an ARINC 565 limit
is exceeded when entering a value and, ARINC limits are enabled, the display shows a warning
message. Refer to Section 6 for details of the factory-defined operating envelope ARINC565.
In the SETUP function, previously stored sets of limits can be selected for use. Each set of limits is
stored under name or number up to a maximum of 10 characters. To enable other sets of limits
or enter additional sets of limits requires the supervisor PIN.
F1
F2
F3
F4
F5
F6
F2
Units
Main Menu
ft kts
ft/min
m km/h
m/min
mbar
mbar/min
Save
Settings
Units
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...