Druck ADTS 505 User Manual
5 - 15
K0260 Issue No. 8
5
T
ABLE
5-3 E
RROR
M
ESSAGES
(
CONTINUED
)
No.
Message
Probable Cause
Action
219
Abort sequence started by user
Abort key pressed on
hand terminal
Wait until system completes abort
sequence
220
"Cannot Modify Altitude
Correction in Pressure Units"
Altitude Correction
Change units in setup to aero units
221
"Changed Aim to ARINC Limit"
Value out of ARINC limit
range
Re-enter value within allowed range
222
Calibration data failed
Incorrect pressure
value entered during
calibration
Check and enter correct pressure
value for reference pressure point
223
"Cannot enter EPR in Aero Units"
EPR selected with aero
units displayed
Change units in setup to pressure
units then select EPR
227
"Changed Aim to Current Limit"
Value out of limit range Re-enter value within allowed range
228
"Limit Overlap"
Value conflicts with
limit ranges
Re-enter value within allowed range
229
"Limit Saved"
Value out of acceptable
range
Re-enter value within allowed range
241
"Ground Request in Leak Mode"
System must be in
control mode
Select control mode
243
"LDK has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress Check custom limit sets not lost
245
"Controller has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress Check/correct calibration
247
"Handset has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress
250
"Vacuum pump processor has
failed"
Internal fault
Return unit to repair depot
251
"Vacuum pump lead is
disconnected"
-
Return unit to repair depot
252
"Vacuum pump transistor is faulty
or drive inhibited"
Internal fault
Return unit to repair depot
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...