6 - 28
Reference and Specification
K0260 Issue No. 8
2.
Use the Up and Down keys to highlight the required aeronautical and pressure units.
3.
Press SAVE SETTINGS to store in non-volatile memory or press MAIN MENU to store in volatile memory
and return to the Main Menu Screen. Press RESTORE DEFAULTS to restore the last set of stored values
and unit selection saved as a default.
Select limits
Enter the PIN to access the SETUP LIMITS
display. Select one of the preset limits of MAX,
STANDARD and CIVIL or define and enable USER
1 limits. An eight character name/code can
replace USER 1 title.
Default limits
This set of limits, identified as: CIVIL are enabled
when the ADTS 505 is switched on:
MIN ALT
-1000 ft
MAX ALT
50,000 ft
MIN CAS
0.0 knots
MAX CAS
450 knots
MAX MACH
1.0 Mach
MAX ROC
6,000 ft/min
ARINC LIMITS
OFF
ALT CORRECTION
0 ft
MIN Ps
115.972 mbar (2.7 inHg)
MAX Ps
1050.406 mbar (38.1000 inHg)
MIN Qc
0.0 mbar
MAX Qc
368.01 mbar (10.8673 inHg)
MAX Rate Ps 109.85 mbar/min (10.8127 inHg/min)
MAX Rate Qc 109.85 mbar/min (10.8127 inHg/min)
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...