Druck ADTS 505 User Manual
6 - 7
K0260 Issue No. 8
6
HELP
•
The HELP key provides information on each key function. The help message generally gives
associated functions and ways of changing the use of each key. When the display shows the
main pressure display, the user can get help on any of the keys on the key-pad. To do this:
•
Press the required key for information, then press HELP.
•
Press
CLEAR/QUIT
to exit the help system.
ENTER
•
Press
ENTER
to complete a numeric entry.
CLEAR/QUIT + ENTER (ABORT)
•
When
CLEAR/QUIT
and
ENTER
are pressed simultaneously, the ADTS 505 restarts from the
power-up sequence.
•
When these keys are pressed all output valves immediately close, isolating the aircraft
systems.
•
The
ABORT
function should only be used as a last resort:
•
When a restart is required after the system has shutdown displaying an error message.
•
As an emergency measure.
•
If the system is restarted by pressing
ABORT
, it vents to atmosphere during the power-up
sequence. The aircraft systems remain at the original pressures. At a later stage in the
power-up sequence, the aircraft system pressures are sampled and the Ps (static) and Pt
(pitot) channels reconnected with a small pressure loss.
Summary of Contents for Druck ADTS 505
Page 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Page 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Page 54: ...3 26 Operation K0260 Issue No 8 ...
Page 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Page 56: ...3 28 Operation K0260 Issue No 8 ...
Page 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Page 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Page 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...