6 - 26
Reference and Specification
K0260 Issue No. 8
Ground
Selected in Control Mode
1.
Use the GROUND touch screen key to select go-to-ground. The displayed channel aims show
GROUND, the controller takes all channels to ground pressures and Qc to zero as shown above.
Note:
At any time during this operation, pressing Leak/Measure stops the operation.
2.
The display shows At Ground Pressures (see above). Pressing “OK” cancels the yellow message box but
leaves the system “At Ground”. Pressing Leak/Measure returns the system to Leak/Measure mode.